NOTICE:
In case of a display range overflow in totalizing, the program displays error
<5-FULL>. The user must either take the load off the pan and press
key in order to complete totalizing or replace the load with another one (of
lower mass) not resulting in display range overflow.
Function deactivation:
Press F key
twice.
15.
ADJUSTMENT
Only for verified scales
In order to ensure perfect weighing accuracy it is required to introduce into
the scale memory a factor correcting the indications – it is so called
adjustment. Adjustment should be performed when the user starts weighing
or when dynamic change of temperature occurs. Before starting the
procedure, loads should be removed from the pan.
15.1. Adjustment procedure
Procedure:
• Following point 11 of this manual enter