Start-up, Scale placing – RADWAG PUE C indicator User Manual
Page 10

EXI-01-04-06-14-GB
- 10 -
3.
START-UP
Please read this manual carefully prior the start-up. Following the
instruction, use the device as intended.
3.1. Scale placing
• The scale should be unpacked outside the hazardous area.
• Before moving the scale to the place of operation prepare a cable
for earthing the device against electrostatic charges.
NOTICE!!
In order to protect the indicator against gathering electrostatic
charge it is necessary to earth/zero its housing.
Connect the earthing lead/wire to cable connection points
marked with „
” . See the manual for details about earthing.
• Situate scales and an indicator in hazardous area and connect
earthing/zeroing.
Functional
grounding mark
Protective sticker
Verification sticker
Rating plate of the scale