Gb 2. safety – WIKA CPH62I0 User Manual

Page 7

background image

7

WIKA operating instructions, model CPH62I0

11221801.02 02/2011 GB/D/F/E

GB

2. Safety

Special operating conditions require further appropriate knowledge, e.g. of aggressive

media.

2.3 Additional safety instructions for instruments with ATEX approval

WARNING!

Non-observance of these instructions and their contents may result in

the loss of explosion protection.

The CPH62I0 hand-held pressure indicator is used for measuring

absolute and/or gauge pressures. In addition, the 2-channel version

can be used to measure differential pressure. The measurement is

carried out by an external pressure sensor.

Battery operation:

Only the usage of approved 9 V batteries allowed!

Batteries may only be changed outside of the hazardous area!

Approved batteries are:

Battery type Manufacturer Battery name
6F22

GB

GREENCELL, 9 V (1604G)

6LF22

or

6LR61

GP

SUPER Alkaline, 9 V (1604A)

Duracell

DURACELL PLUS, Alkaline, 9 V

Varta

powerone alkaline, 9 V (No. 4122)

Varta

INDUSTRIAL, Alkaline, 9V (No. 4022)

Ex-Zone operation:

Within Ex-Zones the accompanying leather case must be used!

Serial interface:

The serial interface may not be used within hazardous areas.

Connection and operation are only permitted to be used outside of

hazardous areas!

Only the WIKA interface converter may be used!

Pressure sensor:

You must only use model CPT62I0 pressure sensors! Usage of

other sensors may result in damage to the sensor and instrument.

When operating with two model CPT62I0 pressure sensors, be

careful that the sensors are not screwed into, nor are in contact

with, surfaces with differing electrical potentials.