3. Specifications

Specifications

CPT62I0 reference pressure sensor

Pressure connection

1)

G½ B; {flush diaphragm (G1 for 0.1 up to 1.6 bar) or various connection adapters

on request}

Material

Wetted parts

Stainless steel or Elgiloy®, (> 25 bar with NBR seal)

flush diaphragm version: stainless steel {Hastelloy C4}; O-Ring: NBR

3)

{FKM/FPM

or EPDM}

Internal transmission fluid

Synthetic oil, (only for pressure ranges up to 16 bar or flush diaphragm)

{Halocarbon oil for oxygen applications}; {Listed by FDA for food industry}

One year stability

0.2 % of span at reference conditions

Permissible

Medium temperature

1)

°C

-20 ... +50 (T4)

Ambient temperature

°C

-20 ... +50 (T4)

Storage temperature

°C

-40 ... +80

Protection class

IP 67 (Sensor) / IP 54 (Plug)

Compensated range

°C

0 ... 70

Temperature coefficients

Mean TC of zero signal

0.2 % / 10 K (< 0.4 for pressure ranges < 250 mbar)

Mean TC of span

0.2 % / 10 K

Supply circiut

Max. voltage

DC V

Ui = 10,4

Max. strength of current

mA

Ii = 93

Max. power

mW

Pi = 500

Max. effective internal capacitance nF

Ci = 600

Max. effective internal inductance

Li negligible

CE conformity

Pressure equipment directive

97/23/EC

EMC directive

2004/108/EC, EN 61326 Emission (Group 1, Class B) and Immunity (portable

equipment)

ATEX directive

94/9/EC, Category 2G, Ignition protection type Ex ib IIC T4

Connection to CPH62I0

via 1 m cable (plug & play); on request: up to 5 m

Weight

g

approx. 220

1) In an oxygen version, a flush diaphragm model is not available. In an oxygen version model CPT62I0 is only available in gauge pressure ranges ≥ 0.25

bar with media temperatures between -10 … +50 °C and using stainless steel or Elgiloy® wetted parts.

2) For the intrinsically safe version, the use of the interface and analogue output is not permitted within an explosive atmosphere.

3) O-ring made of FKM/FPM or EPDM for flush diaphragm with integrated cooling element.

{}

Items in curved brackets are optional extras for additional price.

For further specifications see WIKA data sheet CT 11.02 and the order documentation.

WARNING!

For further important safety instructions for the operation in hazardous

areas see chapter “2.3 Additional safety instructions for instruments

with ATEX approval”.