6. Commissioning, operation

Menu Param.

Setting

Meaning

´MENU´

or

SEt

ConF

Set Configuration: Generic settings

Unit

mbar, bar, ...

Unit: Unit of display

*, **

SL

oFF / on

Sea-level correction: on or off

*, **

Alti

-200 ... +9999

Altitude: Input of altitude above sea-level [m] (only if on)

*, **

rAtE

Rate: Measuring rate (see chapter 6.3.3)

*

Slo

Slow: Measuring rate (4 Hz filtered, low power consumption)

*

FASt

Fast: Measuring rate, filtered (> 1000 Hz)

*

P.dEt

Peak detection: fast measuring rate, unfildered (> 1000 Hz)

*

t.AVG

1 ... 120

Averaging interval in seconds, used by the averaging function off

oFF

Averaging function deactivated

P.oFF

1 ... 120

Auto Power-Off delay in minutes. Device will automatically switch

itself off if, during this interval, no key is pressed, or there is no

serial communication.

oFF

Auto Power-Off function inactive (continuous operation)

Out

oFF

Function of the output: No output function, lowest power consump-

tion

SEr

Output is serial interface

dAC

Output is analogue output 0 ... 1 V

Adr.

01, 11 ... 91

Communications address of the interface

dAC.

CH1, CH2 or

CH DIF

Choice of the input to be the source for the analogue output (if Out

= dAC)

dAC.0

eg. -5.00 ...

+5.00 mbar

Enter desired value at which the analogue output potential should

be 0 V (if Out = dAC)

dAC.1

eg. -5.00 ...

+5.00 mbar

Enter desired value at which the analogue output potential should

be 1 V (if Out = dAC)

SEt

CAL

Set Calibration: Adjustment of sensors

OFS.1

Sensordep.,

e.g. -5.00 ...

+5.00 mbar

The offset of sensor 1 will be displaced by this value to compen-

sate for deviations in the probe or in the measuring device.

OFF

Zero displacement inactive (=0.0 °)

SCL.1

eg. -5.00 ...

+5.00 mbar

The measuring scale of sensor 1 will be changed by this factor

[%] to compensate deviations of probe or measuring device.

OFF

Scale correction factor inactive (=0.000)

OFS.2

Sensordep.,

e.g. -5.00 ...

+5.00 mbar

The offset of sensor 2 will be displaced by this value to compen-

sate for deviations in the probe or in the measuring device.

OFF

Zero dispacement deactive (=0.0°)

SCL.2

eg. -5.00 ...

+5.00 mbar

The measuring scale of sensor 2 will be changed by this factor

[%] to compensate deviations of probe or measuring device.

OFF

Scale correction factor inactive (=0.000)