WIKA operating instructions, model CPH62I0

11221801.02 02/2011 GB/D/F/E

GB

6. Commissioning, operation

Menu Param.

Setting

Meaning

´MENU´

or

SEt

AL.

Set Alarm: Alarm function configuration

AL.

[1,2,DIF]

on

Alarm function on

no. So

Alarm function on

oFF

No alarm function

AL.Lo

[1,2,DIF]

Min-Range ...

... AL.Hi

Min alarm limit

(not when AL.oFF, Min-Range is the lower display range of the

connected sensor)

AL.Hi

[1,2,DIF]

AL.Lo ...

... Max-Range

Max alarm limit

(not when AL.oFF, Max-Range is the upper display range of the

connected sensor)

SEt

LoGG

Set Logger: Logger function configuration

*

Func

CYCL

Cyclic: logger function ‘cyclic logger’

*

Stor

Store: logger function ‘discrete value logger’

*

oFF

No logger function

*

CYCL

1 ... 3600

Cycle time for cyclic logger [seconds]

Lo.Po

on / oFF

Low-power logger with very low power consumption (only if cyclic

logger with slow measuring rate)

SEt

CLOC

Set Clock: Setting of real time clock

CLOC

HH:MM

Clock: set time hours:minutes

dAtE

TT.MM

Date: set time day.month

YEAr

YYYY

Year: set time year

* If there is existing data in the logger memory, the menus/param-

eters cannot be accessed! If these need to be adjusted, the logger

memory must first be cleared! (see chapter “6.4 Operation of

logger”.)

** This menu can only be invoked if an appropriate sensor is connect-

ed to connection 1. When using a second sensor on connection 2

then settings are adopted.

6.3 Configuration

To change device settings, press the ´SET/MENU´ key for 2 seconds. This will access

the configuration menu (main display: ´SEt´).

Pressing the ´SET/MENU´ key again scrolls down the menus, pressing the ´TARA´

key scrolls across to the associated parameters, which can then be selected with the

´TARA´ key.

The parameters can be altered using the ´MIN´ or ´MAX´ keys. Pressing the ´SET/

MENU´ key again saves the settings and returns you to the main configuration menu.

The ´STORE/QUIT´ key exits the configuration and returns you to standard measuring

mode.