beautypg.com

Gb 3. specifications 3. specifications, Specifications cph62i0 digital instrument – WIKA CPH62I0 User Manual

Page 11

background image

11

WIKA operating instructions, model CPH62I0

11221801.02 02/2011 GB/D/F/E

GB

3. Specifications

3. Specifications

Specifications

CPH62I0 hand-held indicator (complete measuring chain)

Input

1 input for CPH62I0-S1

2 inputs for CPH62I0-S2

Pressure range

1)

bar

0.1

0.16 0.25 0.4

0.6

1.0

1.6

2.5

4.0

6.0

10

Over pressure safety

bar

1

1.5

2

2

4

5

10

10

17

35

35

Burst pressure

bar

2

2

2.4

2.4

4.8

6

12

12

20.5

42

42

Resolution (max.)

depending on pressure range

Pressure range

1)

bar

16

25

40

60

100

160

250

400

600

1000

Over pressure safety

bar

80

50

80

120

200

320

500

800

1200 1500

Burst pressure

bar

96

250

400

550

800

1000 1200 1700 2400 3000

Resolution (max.)

depending on pressure range

Type of pressure

gauge {absolute 0.25 up to 16 bar & vacuum on request}

Accuracy (measurement chain)

0.2 % FS ± 1 digit at reference temperature of 20 °C; optional: 0.1 %

Display

Large LCD display for display of 2 values and additional information

Range of display

max. -19999 up to 19999, depending on sensors used

Units

can be selected: bar, mbar, psi, Pa, kPa, MPa, mmHg and inHg (dependent on

range)

Specifications

CPH62I0 digital instrument

Functions via key press

Min-, Max-memory, Hold, Tare, Offset-correction, Logger (Start/Stop)

Functions via Menu

Min-, Max-alarm (visual), Sealevel (barom. air pressure), Power-Off-Function,

measurement rate: 4/sec (“slow“); > 1000/sec (“fast“); > 1000/sec unfiltered (“peak-

detect“) [via “peak-detect“ and min-/max-memory: pressure peaks of 1.5 msec can

be detected]

Data logger

- discrete value logger: up to 99 recordings incl. time via key press

- cyclic logger: automatic recording up to 10,000 values incl. time

- cycle time: selectable from 1 ... 3600 seconds

Interface (serial)

2)

USB or RS-232 interface via special interface cable

Analogue output

2)

0 ... 1 V; configurable (alternative instead of serial interface, selectable via menu)

Power supply

9 V-zinc-carbon battery, alternative: 9 V rechargeable battery or mains supply

Power consumption

cycle time slow: < 1.6 mA, fast: < 7.0 mA, Low-Power-Logger-Fct.: < 0.3 mA

Permissible

Working temperature

°C

-10 ... +50

Air humidity

% r. H. 0 ... 95 relative humidity without moisture condensation

Storage temperature

°C

-20 ... +70

Housing

Impact-resistant ABS, membrane keyboard, transparent panels with protective leather

case

Weight

g

approx. 160

Connection values

Max. voltage

DC V

Uo = 10,38

Max. strength of current

mA

Io = 93

Max. power

mW

Po = 240

Max. effective internal capacitance nF

Co = 1240

Max. effective internal inductance

Lo negligible

CE conformity

EMC directive

2004/108/EC, EN 61326 Emission (Group 1, Class B) and Immunity (portable

equipment)

ATEX directive

94/9/EC, Category 2G, Ignition protection type Ex ib IIC T4

1) In an oxygen version, a flush diaphragm model is not available. In an oxygen version model CPT62I0 is only available in gauge pressure ranges ≥ 0.25

bar with media temperatures between -10 … +50 °C and using stainless steel or Elgiloy® wetted parts.

2) For the intrinsically safe version, the use of the interface and analogue output is not permitted within an explosive atmosphere.

3) O-ring made of FKM/FPM or EPDM for flush diaphragm with integrated cooling element.

{}

Items in curved brackets are optional extras for additional price.