6. Commissioning, operation

6.3.1 (Unit) Choice of pressure units

Depending on the pressure range, the pressure value can be displayed in any one of the

following units: mbar, bar, Pa, kPa, MPa, mmHg, inHg or psi.

6.3.2 (SL) Sea-level correction for absolute pressure sensors

If an absolute pressure sensor is connected to the instrument, then the device displays

the true absolute pressure measured by the sensor. This does not necessarily equate to

the values given by weather stations, which give the pressure at sea-level. Usually the

sensor is situated above sea-level and therefore, if the value at sea-level (zero) is to be

measured, the pressure loss resulting from the elevation above sea-level must be consid-

ered! To correct the displayed value activate the “Sea-Level-Function“ (SL). (Setting is

only possible, if an abs. pressure sensor is connected to sensor connection 1.)

Then enter the sensor‘s height above sea-level in metres (Alti). When activated, the

display shows the SL-arrow and the device displays the absolute pressure value relative

to sea-level.

When two absolute pressure sensors are connected, the sea level

function for both corresponds to the setting of pressure sensor 1.

6.3.3 (rAtE) Choice of measuring rates: “rAtE-Slo, -P.dEt, -FASt“

Three options of measuring rate are supported. Two of them work with high measur-

ing frequencies of over 1000 measurements per second. Whichever one of these was

selected during configuration (see above), this will be displayed in the lower display:

“P.dEt“ or “FASt“.

6.3.3.1 rAtE-Slo = standard measuring

Measuring rate 4 Hz, averaging and filter functions are active.

Application: Measurement of slowly changing or static pressures, e.g. leak testing,

atmospheric pressure, highest accuracy, high noise immunity (EMI and unstable

measuring signals), low power consumption.

6.3.3.2 rAtE-P.dEt: Peak detection

Measuring rate > 1000 Hz, the value is displayed unfiltered.

Application with logger function: Measurement of pressure spikes or fluctuating

pressures with a resolution of < 1 ms. The cyclic logger function records the arithmetic

mean value, the highest and the lowest peak during the chosen time interval.

Higher power consumption, measuring is sensitive to noise (EMI,..).