6. Commissioning, operation

clear all data points

clear nothing (cancel menu)

clear the last data point

Use the ´MIN´ or the ´MAX´ key to display the required option. The ´STORE/QUIT´ key

selects that choice.

If the logger memory is full, the display will show:

6.4.1 “Func-Stor“: Storing discrete measurements

In this mode, each time the ´Store/Quit´ key is pressed a measurement and its time

stamp will be recorded. The data recorded can be viewed either on the display (an

additional menu item “REAd-LoGG“ is displayed when scrolling through the configura-

tion menu) or through the serial interface and a PC running the GSoft software.

Max. number of measurements: 99

The data recorded includes the following values:

Sensor 1: instantaneous value at that data point

Sensor 1: Min-Peak, Max-Peak since last data point

Sensor 2*: instantaneous value at that data point

Sensor 2*: Min-Peak, Max-Peak since last data point

Instantaneous difference (Sensor 1 - Sensor 2)*: at that data point

Min-Peak, Max-Peak of the difference (Sensor 1 - Sensor 2)*: Min-Peak, Max-Peak

since last data point

Time and date stamp of that data point

* Only for 2-channel version CPH62I0-S2

After each recording “St. XX“ will be displayed for a short time. XX represents the

number of the data point.

If the logger memory already contains data:

When the ´STORE/QUIT´ key is pressed for 2 seconds, the option to clear the logger

memory will be displayed: