7. Maintenance, cleaning and servicing (recalibration) / 8. Faults

7.3 Recalibration
DKD/DAkkS certificate - Certificates:

We recommend that the instrument is regularly recalibrated by the manufacturer, with

time intervals of approx. 12 months. Every factory recalibration includes, additionally, an

extensive free-of-charge check of all system parameters with respect to their compli-

ance with the specification. The basic settings will be corrected if necessary.

8. Faults

Display

Meaning

Measures

Low battery power, device will only

work for a short while longer

Replace battery

Battery flat

Replace battery

Mains operation without battery:

incorrect voltage

Check power supply, change it if

necessary

SEnS

Erro

or Err.9

No sensor connected

Switch off device and connect

sensor

Connected sensor or device faulty

If second sensor available, check

if device is ok. Return defective

device/sensor to manufacturer for

repair

Value significantly out of measuring

range

Check: pressure not within sensor

range?

- - - -

- - - -

Logger data is being read by the

serial interface

When the data transfer is complete

the device will automatically return

to normal measuring display, no

remedy necessary

No display or

strange sym-

bols, device is

not respond-

ing to input

Battery flat

Replace battery

Mains operation without battery:

incorrect voltage or polarity

Check power supply, change it if

necessary

System error

Disconnect battery or power supply,

wait a while, re-connect

Instrument faulty

Return to manufacturer for repair