6. Commissioning, operation

Activating/deactivating ´TARA´ clears the max- and min-memories.

Zero-Point Adjust: (for gauge pressure sensors only) If there is no pressure applied

to the pressure ports the device should display 0. If there is a perma-

nent offset (and the device is being used under steady conditions),

a permanent zero point adjustment can be carried out. To carry out

the adjustment, press ´TARA´ for approx. 5 seconds. (Please note: A

zeropoint adjustment can only be made if the difference between the

display value and the factory calibration value is less than 2 %! E.g.

for a measuring range of 0 ... 25 bar, => zeropoint adjustment up to

0.5 bar is possible). To revert to the factory calibration, press ´TARA´ for

approx. 15 seconds.

A zero-point adjustment can only be carried out if the difference

between the value on display is less than 500 digits!

If a zero-point adjustment has been made, this will be signalled by the

message “nuLL-Corr“ being displayed briefly as the device is turned on.

Differential pressure: Pressing ´SET/MENU´ shows the difference between the

measured values of Channel 1 and Channel 2 (DIF=CH1-CH2) in the

lower display. Pressing it again will hide it (only with the 2-channel

version and when 2 pressure sensors are connected).