beautypg.com

Maintenance and cleaning – WIKA Hydra-Gauge User Manual

Page 12

background image

12

WIKA operating instructions HYDRA-Gauge

9015329.01 08/2013 GB/D/F/E

GB

8. Maintenance and cleaning / 9. Dismounting and disposal

9. Dismounting and disposal

WARNING!

Residual media in dismounted instruments can result in a risk to persons, the

environment and equipment.

Take sufficient precautionary measures.

9.1 Dismounting

Only disconnect the pressure gauge once the system has been depressurised!

9.2 Disposal

Incorrect disposal can put the environment at risk.

Dispose of instrument components and packaging materials in an environmentally

compatible way and in accordance with the country-specific waste disposal regulations.

8. Maintenance and cleaning

8.1 Maintenance

HYDRA-Gauge pressure measuring assemblies are maintenance-free.

The indicator and switching function should be checked once or twice every year.

The instrument must be disconnected from the process to check the indication and switch-

ing function with a pressure testing device.

The HYDRA double diaphragm system prevents contamination of the process medium

even in case of mechanical damage of the wetted diaphragm. In this case the process

medium will penetrate the lateral leakage bore. The instrument must be exchanged as soon

as possible.

For safety reasons repairs of already installed instruments are not possible.

8.2 Cleaning

CAUTION!

The HYDRA-Gauge may only be cleaned by rinsing. Any mechanical

cleaning inside the process connection may damage the separating

diaphragms.

Clean the instrument with a moist cloth.

Electrical connections must not come into contact with moisture.