beautypg.com

WIKA DPT-10 User Manual

Page 81

background image

-

Double chamber housing

l

1 x cable entry M20 x 1.5 (cable: ø 5 … 9 mm),
1 x blind stopper M20 x 1.5; plug M12 x 1 for
the external indicating and adjustment unit
(optional)

or:

l

1 x closing cap ½ NPT, 1 x blind stopper
½ NPT

, plug M12 x 1 for the external indicating

and adjustment unit (optional)

or:

l

1 x plug (depending on the version), 1 x blind
stopper M20 x 1.5; plug M12 x 1 for the
external indicating and adjustment unit (optio-
nal)

Spring-loaded terminals for wire cross-sec-
tion up to

2.5 mm² (AWG 14)

Electromechanical data - version IP 66/IP 68 (1 bar)

Cable entry

-

Single chamber housing

l

1 x IP 68 cable gland M20 x 1.5; 1 x blind
stopper M20 x 1.5

or:

l

1 x closing cap ½ NPT, 1 x blind plug ½ NPT

Connection cable

-

Configuration

four wires, one suspension cable, one breather
capillary, screen braiding, metal foil, mantle

-

Wire cross-section

0.5 mm² (AWG 20)

-

Wire resistance

<

0.036 Ω/m (0.011 Ω/ft)

-

Tensile strength

>

1200 N (270 pounds force)

-

Standard length

5 m (16.4 ft)

-

Max. length

1000 m (3281 ft)

-

Min. bending radius at 25 °C/77 °F

25 mm (0.985 in)

-

Diameter approx.

8 mm (0.315 in)

-

Colour - standard PE

Black

-

Colour - standard PUR

Blue

-

Colour - Ex-version

Blue

Electromechanical data - version IP 66/IP 68 with external electronics

Connection cable between IP 68 instrument and external housing:

-

Configuration

Four wires, screen braiding, inner cover, screen
braiding, outer cover

-

Wire cross-section

0.5 mm² (AWG 20)

-

Standard length

5 m (16.40 ft)

-

Max. length

25 m (82.02 ft)

Differential pressure transmitter DPT10 - metal measuring diaphragm • Profibus PA

81

11 Supplement

37245

-

EN

-100120