Proceed as follows:

1

Edit the % value in the menu item "Min. adjustment" with [OK].

Min. adjustment

0

.00 %

=

0.0000 bar
0.0000 bar

2

Set the requested value with [+] and [->].

3

Confirm with [OK] and edit the requested bar value.

4

Set the requested bar value with [+] and [->].

5

Confirm with [OK] and move to max. adjustment with [->].

For an adjustment with filling, simply enter the actual measured value
indicated at the bottom of the display.

The min. adjustment is finished.

Proceed as follows:

1

Edit the % value in the menu item "Max. adjustment" with [OK].

100

.00 %

=

0

.5000 bar

0.0000 bar

Information:
When the instrument is not yet adjusted, then the display pressure for
100 % corresponds to the nominal measuring range of the sensor (in
the above example 500 mbar).

2

Set the requested value with [->] and [OK].

3

Edit the requested mbar value with [OK].

4

Set the requested value with [+] and [->].

5

Confirm with [OK] and move to the menu overview with [ESC].

For an adjustment with filling, simply enter the actual measured value
indicated at the bottom of the display.

The max. adjustment is finished.

Proceed as follows:

1

Edit the bar value in the menu item "Min. adjustment" with [OK].

Min. adjustment

0

.00 %

DP

-

+

=

0.0000 bar
0.0000 bar

Min. adjustment with
level

Max. adjustment with
level

Min. adjustment with
flow

Differential pressure transmitter DPT10 - metal measuring diaphragm • Profibus PA

49