7. Performing Melt Determinations

Once the analysis method and sample information have been entered into the SMP40 software the
instrument will display the New Measurement screen.

The Capillary tubes melt status is for Information Only. An accurate analysis of the data is
performed after the melt determination has been completed.

1 Place 1-3 prepared melting point capillaries

into the SMP40 sample compartments sample
block.

The external surface of the sample
capillaries must be wiped clean before being
placed into the sample block as chemical
residues may build up and affect the
performance of the device.

2 Press “Start the melt” to initiate the melting

protocol.

14

Current block
temperature

Sample name

Method

information

Capillary

tubes melt
status

Live image
of the
sample
compartment

Figure 6: New Measurement Screen