6.1 Creating a New Method

In the Measurement menu screen touch the “Create a new measurement” section of the screen.

1 Enter a sample name for the samples by

selecting the Sample Name field

2 Enter an appropriate start or plateau

temperature for the sample by selecting the
Start Temperature field.

3 Enter the required temperature ramp rate for

the samples by selecting the Ramp Rate field.

4 Optional sample and method information is

accessed by pressing the “Touch here to enter
additional information” button. If this is not
required press “Continue” and go to step 8.

5 The SMP40 will automatically detect the end

of the melt when the toggle selection field
has an X in the Automatically detect the melt
end field. Removing this check mark allows a
value to be entered into the End Temperature
field.

6 Enter any batch or reference details in the

Batch/Reference alphanumeric entry field.

7 To proceed to the melt determination select

“Go back to main information” and then
select “Continue”.

8 The option of saving this new method as a

preset method will be given. Select “Yes” to
save the method for future use or select
“No” to proceed without saving the method.