Dynalab SMP40 User Manual
Page 13

11
6.1 Creating a New Method
In the Measurement menu screen touch the “Create a new measurement” section of the screen.
1 Enter a sample name for the samples by
selecting the Sample Name field
2 Enter an appropriate start or plateau
temperature for the sample by selecting the
Start Temperature field.
3 Enter the required temperature ramp rate for
the samples by selecting the Ramp Rate field.
4 Optional sample and method information is
accessed by pressing the “Touch here to enter
additional information” button. If this is not
required press “Continue” and go to step 8.
5 The SMP40 will automatically detect the end
of the melt when the toggle selection field
has an X in the Automatically detect the melt
end field. Removing this check mark allows a
value to be entered into the End Temperature
field.
6 Enter any batch or reference details in the
Batch/Reference alphanumeric entry field.
7 To proceed to the melt determination select
“Go back to main information” and then
select “Continue”.
8 The option of saving this new method as a
preset method will be given. Select “Yes” to
save the method for future use or select
“No” to proceed without saving the method.