beautypg.com

4 logchart-ii software, 1 installing logchart-ii, 2 running logchart-ii – NOVUS TagTemp User Manual

Page 2: 3 configuring the logger, 4logchart-ii software

background image

TAGTEMP logger

NOVUS AUTOMATION

2/7

4

LOGCHART-II SOFTWARE

4.1

INSTALLING LOGCHART-II

LogChart II is the software provided with the logger to allow for

configuration and data offload. To install, run the LC_II_Setup.exe

file provided with the CD-ROM.
Note: Be sure your Windows date separator is configured as a slash:

dd/mm/yy or dd/mm/yyyy.

4.2

RUNNING LOGCHART-II

When you open LogChart-II the main window is displayed.

Figure 02 – LogChart-II main window

Next, select the serial port that the communication interface will use

in the “Port” menu.
Check which serial is available. Usually COM2 is free, once the

mouse uses COM1. The chosen port will be chosen by default next

times the LogChart-II is initiated. When a valid port is selected, the

icons below are displayed.

Figure 03 – Icons enabled when the communication port selected is a valid

port

4.3

CONFIGURING THE LOGGER

Make sure the communication interface is connected to the PC port

selected. The interface must be constantly directed towards the front

part of the logger (communication window) at a maximum distance of

1 m. (See Figure 05).

Figure 04 – Infrared communication interface position

The logger must be “waken up” by pressing the Start/Stop button.
After the serial port is selected, select:

The Parameters Configuration screen is displayed. In this screen

the user can define the logger operation mode and also obtains

general information about the device.

Figure 05 – Configuration window

Fields are:
1- Title: In this field, the user identifies the logger by assigning it a

name.
2- General information: Area with information about the logger,

such as Model, Serial Number, Logger Date/Time, PC Date/Time,

firmware version, memory capacity and number of acquisitions

stored in memory, etc.
In this field, time is constantly updated while the communication

between logger and computer is taking place.
3- Readings: Presents a series of parameters that define the

acquisition process.

Interval: Defines the interval between readings: The minimum

interval is (1) second and maximum is (18) hours.
Note: When the type of value logged is maximum or minimum

values, the minimum interval is 10 seconds.
In Daily repetitions the user defines the time that daily logs will

take place.
Estimated time: In this parameter, the logger informs the user

how long it will take to occupy the full memory, in the conditions

set up during configuration.
Start of Readings: Readings can be started in one of five

different modes:

Immediately: start as soon as set up is ready and sent

(OK) to the logger.

Start via Palm: start with a command sent via PDA, which

runs the software Log Chart Palm-OS. For further

information, see PDA User section in this guide.

Date: readings start at predefined date and time.

Via Start/Stop Button: starts and interrupts readings by

pressing the Start/Stop button, in the frontal part of the

logger, for two seconds. See Figure 01.

Setpoint: acquisitions start when a temperature setpoint

is reached. In this option, the setpoint value is defined in

the Channels field, where the Alarm parameter is replaced

by setpoint.

Stop loggings: Options for the end of readings are:

Full Memory: readings can be stored up to the full

memory capacity is reached.

Don’t stop (Wrap around): readings are continuous,

replacing old records with new ones as the number of

readings overpass the memory capacity.

After: the logger stops readings after a certain number of

readings.

Date: Readings stop at user-predefined date and time. In

case the logger memory capacity is reached before the

date defined, readings are stopped.