beautypg.com

Electrical connections, Operation – NOVUS TxBlock-USB Transmitter User Manual

Page 3

background image

TxBlock-USB

NOVUS AUTOMATION

3/4

ELECTRICAL CONNECTIONS
The figures below show the electrical connections required. The

terminals 3, 4, 5 and 6 are dedicated to the sensor connection.

LOAD represents the 4-20 mA current measuring device (indicator,

controller, recorder, etc.).

PT100 2-WIRES
Note
: When the Pt100 2-wires the terminals 3 and 4 must be

interconnected, according to the figure below.
The Pt100 wire length should be less than 30 cm to maintain the

measurement error within specifications (electrical resistance).

Fig. 5 – Transmitter electrical connections (Pt100 2-wires)

PT100 3-WIRES

Fig. 6 – Transmitter electrical connections (Pt100 3-wires)

PT100 4-WIRES

Fig. 7 – Transmitter electrical connections (Pt100 4-wires)

Pt100 / Pt1000 3 and 4 wires: For appropriate cable resistance

compensation they should be equal for all legs. Maximum wire

resistance is 25 Ω per wire leg. Usage of a 3 or 4 wire with conductors

of equal length and gauge is recommended.

THERMOCOUPLES

Fig. 8 – Transmitter electrical connections (Thermocouple)

VOLTAGE (0-50 mV)

Fig. 9 – Transmitter electrical connections (0-50 mV)

OPERATION

The sensor offset can be changed through the TxConfig II software.

The USB cable may be connected to the transmitter without causing

any measurement errors. See item Zero Correction in the chapter

CONFIGURATION of this manual.
The uer must choose the most suitable sensor and range to the

process. The chosen range must not exceed the maximum range of

measurement defined for the sensor and should not be smaller than

the minimum range for the same sensor.
It is important to note that the transmitter accuracy is based on the

maximum range of the sensor used, even when a narrower range is

programmed. Example:

The sensor Pt100 has a max. range of -200 to +650 °C and

accuracy of 0.12 %, thus we will have an error up to 1.7 °C

(0.2 % of 850 °C).

This error may be present even when a narrower range is

configured for the sensor. (For example: 0 to 100 °C.)

Note: When measurements are made at the transmitter, see if the

Pt100 excitation current required by the calibrator is compatible with

the Pt100 excitation current used in the transmitter: 0.8 mA.