A.5.1 configcmd 11 zero-and-span control, A.5.1 – Campbell Scientific EC150 CO2 and H2O Open-Path Gas Analyzer and EC100 Electronics with Optional CSAT3A 3D Sonic Anemometer User Manual
Page 59

Appendix A. EC150 Settings
TABLE A-2. ConfigCmd Values for Setting and Retrieving Settings
ConfigCmd Variable Setting Description (some settings list possible
values for the DestSource variable)
Set
Retrieve
6
106
Pressure Gain
7
107
Temperature Sensor:
•
0 = EC150 Temperature Probe
•
1 = EC155 Sample Cell Thermistor
•
2 = EC155 Sample Cell Thermocouple
•
3 = None (use fixed value)
•
4 = Auto-Select
8
108
Fixed Temperature Value
9
109
Unprompted Output Mode:
•
0 = Disable
•
1 =USB
•
2 = RS-485
10
110
RS-485 Baud Rate
11
111
Span/Zero Control:
•
0 = Inactive
•
1 = Zero
•
2 = Span CO
2
•
3 = Span H
2
ConfigCmd 11 Zero-and-Span Control.)
12
112
CO
2
Span Concentration
13
113
H
2
O Span Dewpoint Temperature
14
114
CO
2
Zero
15
115
CO
2
Span
16
116
H
2
O Zero
17
117
H
2
O Span
18 or 218
118
Heater Voltage (0 to 4.5375V, −1 = Off)
19
119
Reserved
20
120
Analog Output Enable:
•
0 = Disable
•
1 = Enable
21
121
PowerDown:
•
0 = Gas Head On
•
1 = Gas Head Off
99
N/A
Save Settings to EEPROM memory
A.5.1 ConfigCmd 11 Zero-and-Span Control
To perform zeroing of CO
2
and H
2
O, ConfigCmd 11 is set to
1. After the
EC150 completes the zero, it will write the value to
–1. The datalogger can
poll this value or simply wait for a period of time to allow the zeroing to
A-11