TABLE 8-3. Diagnostic Flags and Suggested Actions

Flag

Number

Flag Name

Comments

18

CO

2

Signal

Strength

Set if the ratio of the CO

2

measurement

and the CO

2

reference signals are

outside prescribed limits. It can be

turned on when the measurement path is

obstructed by insects, dust, precipitation,

condensation etc. If it persists, consult

with a Campbell Scientific application

engineer.

19

H

2

O Signal

Strength

Set if the ratio of the H

2

O measurement

and the H

2

O reference signals is outside

prescribed limits. It can be turned on

when the measurement path is

obstructed by insects, dust, precipitation,

condensation etc. If it persists, consult

with a Campbell Scientific application

engineer.

20

Calibration Error Set if there is a problem reading or

writing into the analyzer’s head

memory. Power off the EC100 and

reconnect the head. If the flag persists,

consult with a Campbell Scientific

application engineer. For more

information on the head memory refer to

Section 4.3, Gas Head Memory.

21

Heater Ctrl Off

22

Not Used