Appendix A. EC150 Settings
The instruction syntax is:
EC100Configure(Result,SDMAddress,ConfigCmd,DestSource)
Result is a variable that contains a value indicating the success or failure of the
command. A result code of 0 means that the command was successfully
executed. If reading a setting, 0 in the result code means that the value in the
DestSource variable is the value the desired setting has in the EC150. When
writing a setting, if the result code is 0, the value and setting were compatible,
but the value was not changed because it contained the same value that was
sent. A return code of 1 from the set operation means that the value was valid,
different, set and acknowledged. This allows CRBasic code to control whether
or not to save the settings. NAN (not a number) indicates that the setting was
not changed or acknowledged or a signature failure occurred.
SDMAddress defines the address of the EC150 to configure. Valid SDM
addresses are 0 through 14. Address 15 is reserved for the
SDMTrigger()
instruction.
ConfigCmd is a variable that indicates whether to get or set a setting. The
options are listed in TABLE A-2.
DestSource
is a variable that will contain the value to read when getting a
setting, or that will contain the value to send when writing a setting to the
EC150.
TABLE A-2. ConfigCmd Values for Setting and Retrieving Settings
ConfigCmd Variable Setting Description (some settings list possible
values for the DestSource variable)
Set
Retrieve
0
100
Bandwidth:
•
5 = 5 Hz
•
10 = 10 Hz
•
12 = 12.5 Hz
•
20 = 20 Hz
•
25 = 25 Hz
1
101
Unprompted Output:
•
10 = 10 Hz
•
25 = 25 Hz
•
50 = 50 Hz
2
102
Pressure Sensor:
•
0 = EC100 Basic
•
1 = User-supplied
•
2 = EC100 Enhanced
•
3 = None (use fixed value)
3
103
Differential Pressure:
•
0 = Disable
•
1 = Enable
4
104
Fixed Pressure Value
5
105
Pressure Offset