8. Maintenance and Troubleshooting

EC150 operation requires six maintenance tasks:

Routine site maintenance

Wick maintenance

Analyzer window cleaning

Zero and span

Replacing the analyzer desiccant/scrubber bottles

Factory recalibration

8.1 Routine Site Maintenance

The following items should be examined periodically:

Check the humidity indicator card in the EC100 enclosure. If the highest

dot has turned pink, replace the desiccant bags. Replacement desiccant

bags may be purchased as pn 6714.

Make sure the Power and Gas LED status lights on the EC100 panel are

green. If not, check the individual diagnostic bits for the specific fault.

See TABLE 8-2, Diagnostic Flags of Sonic Status LED, and Section 8.7.3,

LED Status Lights, for more information.

8.2 Gas Analyzer Wicks

The windows of the EC150 gas analyzer are polished and slanted at an angle to

prevent water from collecting on their surfaces. However, due to increased

surface tension at the interface with the snout, water can pool at the edges and

partially block the optical path and attenuate the signal. To minimize the

occurrence of such events and the resulting data loss, consider using the wicks

listed in TABLE 8-1. The weave of the wicking fabric promotes capillary

action that wicks the water away from the edge of the windows. The seam and

the straight edge of the wicks are permeated with a rubberized compound to

prevent them from shifting during operation.

Proper installation of the wicks is critical. They should not block or encroach

on the optical path. Before installation, record signal strengths for both H

2

O

and CO

2

. Following installation, repeat testing of signal strength and check

that these values are unchanged.

TABLE 8-1. Rain Wick Replacement Parts

Wick

Campbell Scientific, Inc.

Part Number

Campbell Scientific, Inc.

Description

Bottom wick

28652

Fab EC150/IRGASON rain

wick bottom

Top wick

28653

Fab EC150/IRGASON rain

wick top


The top wick has a short seam which must be aligned with the short side of the

top snout. The angled edge of the wick must closely follow the edge of the

window without encroaching on the optical path. See FIGURE 8-1 for

guidance on proper positioning of the wick.