beautypg.com

8 theory of operation, Theory of operation, Ep p – Campbell Scientific EC150 CO2 and H2O Open-Path Gas Analyzer and EC100 Electronics with Optional CSAT3A 3D Sonic Anemometer User Manual

Page 16

background image

EC150 CO

2

/H

2

O Open-Path Gas Analyzer

The algorithm uses the following environmental parameters to control the

heater:

Analyzer body temperature, measured inside the source housing

(heater control does not allow the body temperature to drop below

ambient air temperature)

Ambient relative humidity (in humidity greater than 80% heaters will

try to maintain internal temperature 2 degrees warmer than ambient)

CO

2

signal level (1 min average CO

2

signal level; below 0.7 will

cause the heater to turn on maximum power until the signals recover)

Average slope of the CO

2

signal level over 1 min

Standard deviation of the CO

2

signal over 1 min

4.8 Theory of Operation

The EC150 is a non-dispersive mid-infrared absorption analyzer. Infrared

radiation is generated in the upper arm of the analyzer head before propagating

along a 15.0 cm (5.9 in) optical path as shown in FIGURE 5-1. Chemical

species located within the optical beam will absorb radiation at characteristic

frequencies. A mercury cadmium telluride (MCT) detector in the lower arm of

the gas analyzer measures the decrease in radiation intensity due to absorption,

which can then be related to analyte concentration using the Beer-Lambert

Law:

cl

o

e

P

P

ε

=

where:

P is irradiance after passing through the optical path

P

o

is initial irradiance, ε is molar absorptivity, c is analyte

concentration, and

l is path length.

In the EC150, radiation is generated by applying constant power to a tungsten

lamp which acts as a 2200 K broadband radiation source. Specific

wavelengths are then selected using interference filters located on a spinning

chopper wheel. For CO

2

measurements, light with a wavelength of 4.3 µm is

selected as that corresponds to the asymmetric stretching vibrational band of

the CO

2

molecule. For H

2

O, the symmetric stretching vibration band is 2.7

µm.

The EC150 is a dual wavelength, single-beam analyzer. This design eliminates

the need for a separate reference cell and detector. Instead, the initial intensity

of the radiation is calculated by measuring the intensity of nearby, non-

absorbing wavelengths (4.0 µm for CO

2

and 2.3 µm for H

2

O). These

measurements mitigate measurement inaccuracy that may arise from source or

detector aging, as well as for low-level window contamination. For window

contamination that reduces the signal strength below 0.8, windows should be

cleaned as described in Section 8.3, Cleaning Analyzer Windows.

The chopper wheel spins at a rate of 50 revolutions per second and the detector

is measured 1024 times per revolution, resulting in a detector sampling rate of

51.2 kHz. The detector is maintained at −40°C using a three-stage

thermoelectric cooler and is coupled to a low noise pre-amp module.

6