beautypg.com

Operating instructions – CDN IN428 - Infrared Thermometer EN User Manual

Page 2

background image

Operating Instructions

A. Temperature Scale

To select temperature reading in Fahrenheit

or Celsius:
1. Press the SCAN button to turn the

thermometer on.

2. Press the MODE button four times.

The

°F or °C symbol flashes on the display.

3. Press the SCAN button to change the scale.

B. Infrared Thermometer

1.

Infrared Scanning
a.

Distance:Spot = 1:1

For example, if the surface area being

measured is 10" in diameter, then the

thermometer must be within 10" of the target

for an accurate reading.

b. Aim the thermometer at the target and press

the SCAN button to display the surface

temperature.

2.

Minimum Mode
a. Press the SCAN button to turn the

thermometer on.

b. Press the MODE button once. The

MIN icon

flashes on the display.

c. Press and hold the SCAN button to confirm

the Minimum Mode and display the lowest

temperature among multiple targets.

3.

Maximum Mode
a. Press the SCAN button to turn the

thermometer on.

b. Press the MODE button twice. The

MAX

icon flashes on the display.

c. Press and hold the SCAN button to confirm

the Maximum Mode and display the highest

temperature among multiple targets.

4.

Lock Mode

This is particularly useful for continuous

temperature monitoring.
a. Press the SCAN button to turn the

thermometer on.

b. Press the MODE button three times.

The

LOCK icon flashes on the display.

c. Press the SCAN button to confirm the Lock

Mode. The IN428 will continuously display

the temperature for up to 60 minutes or until

the SCAN button is pressed again.

5.

Emissivity

Everything gives off a certain amount of

radiation. Emissivity is the measure of this

thermal radiation. The infrared thermometer is

supplied with a default emissivity of 0.95, which

standard for most uses. The emissivity of the

thermometer can be changed from 0.05 (5E) to

1 (100E).

Only experienced personnel should

attempt to make changes.

For information

relating to the emissivity of specific materials,

please contact CDN.
a. Press the SCAN button to turn the

thermometer on.

b. Press the MODE button five times to enter

Emissivity Mode.

95E flashes on the display.

c. Press the SCAN button to adjust the

emissivity value in 0.01 (1E) increments.

d. Press the MODE button again to exit

Emissivity Mode.

Note:

Infrared thermometers are not recommended

for use in measuring the temperature of shiny or

polished metals.

6.

Error Messages

The IN428 incorporates visual diagnostic

messages as follows:
a.

HI or LO is displayed when the temperature

being measured is outside of the range of the

thermometer.
1. HI indicates that the temperature

is higher than 428°F/220°C.

2. LO indicates that the temperature

is lower than -67°F/–55°C.

b. Allow a minimum 30 minutes for the

thermometer to stabilize to the working/

room temperature.
1. ER2 is displayed when the thermometer

is exposed to rapid changes in the

ambient temperature.

2. ER3 is displayed when the ambient

temperature exceeds 14°F/-10°C OR

+122°F/+50°C.

c. For all other error messages

it is necessary to reset the IN428.
1. Wait for the thermometer to power off.
2. Remove the battery and wait for a

minimum of one minute.

3. Reinstall the battery (see

Battery

Installation).

4. Press the SCAN button to turn the

thermometer on.

5. If the error message remains, please

contact CDN for further assistance.

EMC/RFI

Readings may be affected if the unit is operated within a radio

frequency electromagnetic field strength of approximately

3 volts per meter, but the performance of the instrument will

not be permanently affected.