Vaisala WMT52 User Manual
Page 98

User’s Guide ______________________________________________________________________
96 ___________________________________________________________________M210925EN-B
Command format in ASCII and NMEA 0183: aWU
Command format in SDI-12: aXWU!
The response in ASCII and NMEA 0183:
aWU,R=[R],I=[I],A=[A],G=[G,U=[U],D=[D],N=[N],F=[F]
lf>
The response in SDI-12:
aXWU,R=[R],I=[I],A=[A],G=[G],U=[U],D=[D],N=[N],F=[F]
where [R][I][A][G][U][D][N] are the setting fields, see the following
sections.
Example (ASCII and NMEA 0183, device address 0):
0WU
0WU,R=01001000&00100100,I=60,A=10,G=1,U=N,D=
90,N=W,F=4
Example (SDI-12, device address 0):
0XWU!0XWU,R=11111100&01001000,I=10,A=3,
G=1,U=M,D=0,N=W,F=4
where
a
=
Device address
WU
=
Wind sensor settings command in ASCII and NMEA
0183
XWU
=
Wind sensor settings command in SDI-12
=
Command terminator in ASCII and NMEA 0183
!
=
Command terminator in SDI-12