Chapter 10 troubleshooting, Chapter 10, Table 9 – Vaisala WMT52 User Manual
Page 111: Data validation, Troubleshooting

Chapter 10 __________________________________________________________ Troubleshooting
VAISALA______________________________________________________________________ 109
CHAPTER 10
TROUBLESHOOTING
This chapter describes common problems, their probable causes and
remedies, and includes contact information for technical support.
Table 9
Data Validation
Problem
Interpretation
Action
Wind measurement failure. Both
the speed and direction units are
replaced by a # sign or the data
values are irrelevant.
Blockage (trash, leaves,
branches, bird, snow, ice)
between the wind transducers.
Incorrect
terminal program.
Remove the blockage, and
check that the wind transducers
are not damaged.
If the blockage is ice or snow, it
will melt after some time if
heating is enabled. Time for
clearance depends on the
severity of the weather event. If
birds are causing the blockage,
consider using the bird spike kit.
In ASCII and NMEA protocols
both
after each command. Check that
your terminal program sends
both when pressing enter.
Note: The direction unit is # for
the wind speeds less than 0.05
m/s.