Error Messaging/Text Messages Table
Text message identifier (in
NMEA 0183 v3.0 protocol only)
Text Message
Interpretation and action
01
Unable to measure error
The requested parameters are
not activated in the message,
see
,
and check the parameter
selection fields.
02
Sync/address error
The device address in the
beginning of the command is
invalid. Ask the device address
with the ?! (SDI-12) or ?
(ASCII and NMEA) command
and retype the command with
the correct address.
03
Unknown cmd error
The command is not supported,
use the correct command format,
see
.
04
Profile reset
Checksum error in configuration
settings during power-up.
Factory settings used instead.
05
Factory reset
Checksum error in calibration
settings during power-up.
Factory settings used instead.
06
Version reset
New software version in use.
07
Start-up
Software reset. Program starts
from the beginning.
08
Use chksum xxx
Given checksum not correct for
the command. Use the proposed
checksum.
09
Measurement reset
The ongoing measurement of all
the sensors interrupted and
started from the beginning.
NOTE
When WMT52 is used on an RS-485 bus with other polled devices, the
error messaging feature shall always be disabled. This is done by the
following command: 0SU,S=N