Calculated Variables

Dewpoint converted to dewpoint in atmospheric pressure (T

d/f

atm).

Operating Environment

Temperature

-10 ... + 60 °C (+14 ... +140 °F)

Operating Pressure

0 ... 20 bar

a

(290 psi

a

)

Relative humidity

0 … 100 %RH

Sample flow rate

no effect on measurement accuracy

Measured gases

non-corrosive gases

Inputs and Outputs

Analog output (scalable)

4...20 mA, 2-wire

Resolution for current output

0.002 mA

Typical temperature dependence

0.005% of full scale / °C

Connector

4-pin M8 (IEC 60947-5-2)


LED indication available for defined dewpoint limit/error state indication

RS485 serial line for service use

General

Sensor Vaisala

HUMICAP®180R

Recommended calibration interval

2 years

Mechanical connection

G 1/2" ISO

Operating voltage

10 .... 28 VDC

External load

max 100 Ω for supply voltages 10 ...20 VDC

max 500 Ω for supply voltages 20 ... 28 VDC

Weight

65 g (2.3 oz)

Housing material

PPS + 40% GF

Housing classification

IP65 (NEMA 4)

Storage temperature range

-40 ... + 80 °C (-40 ... +176 °F)

Start-up time

3 s


Complies with EMC standard EN61326-1, Electrical equipment for
measurement, control and laboratory use - EMC requirements; Industrial
environment.