View User Adjustment Parameters

Use the L command to view the current user adjustment parameters that
are affected by the CRH, RHCAL, CT, and LI commands. This
command is useful for checking the currently applied correction.

L

The output values are as follows:

- Cp offset and gain: capacitance, adjusted using the CRH command.

- RH offset and gain: adjusted using the RHCAL command.

- T offset and gain: adjusted using the CT command.

Example (shows default values, no user adjustment done):

>l
Cp offset : 0
Cp gain : 1
RH offset : 0
RH gain : 1
T offset : 0
T gain : 1
>

Set User Adjustment Parameters

Use the LI command to set the user adjustment parameters directly. This
command is useful for restoring the desired adjustments, and removing
the effects of user adjustment without having to perform the factory reset.

LI

After entering the command, you will be prompted to enter new values
one at a time. Press ENTER without giving a value to move to the next
parameter.

Example (shows default values):

>li
Cp offset : 0 ?
Cp gain : 1 ?
RH offset : 0 ?
RH gain : 1 ?
T offset : 0 ?
T gain : 1 ?
>