Chapter 6 ___________________________________________________________ Troubleshooting

VAISALA _______________________________________________________________________ 65

Error Codes

The transmitter has error codes that can be displayed on the serial line
using the ERRS command. See section Display Currently Active Errors
on page 41.

Table 8

Error Codes

Error Code Error Message

Possible Cause(s)

E1 Temperature

sensor

open circuit/not calibrated

- Temperature sensor damaged
- Temperature sensor short circuit
- Temperature sensor missing

E2

Humidity sensor open
circuit/not calibrated

- Condensation on humidity sensor
- Humidity sensor damaged
- Humidity sensor short circuit
- Humidity sensor missing

E5

Capasitance out of range - Condensation on humidity sensor

- Humidity sensor damaged

E6

Program checksum error Internal error

E7 Parameter

storage

checksum error

Internal error

E9 Sensor

coefficient

checksum error

Internal error

E10 Current

settings

checksum error

Internal error

E11 Factory

settings

checksum error

Internal error

E12 Temperature

sensor

values out of range

- Temperature sensor damaged
- Temperature sensor missing

E14

MeasTask not calculating
(ATEST)

This error may occurr when the
analog output is in test mode. See
page Test Analog Output on page
42.

E16

Default settings not found Internal error

E17

RH reading too high

Internal error

E18

RH reading too low

Internal error

E20

Log corrupted

Internal error