Error codes – Vaisala DMT132 User Manual
Page 67

Chapter 6 ___________________________________________________________ Troubleshooting
VAISALA _______________________________________________________________________ 65
Error Codes
The transmitter has error codes that can be displayed on the serial line
using the ERRS command. See section Display Currently Active Errors
on page 41.
Table 8
Error Codes
Error Code Error Message
Possible Cause(s)
E1 Temperature
sensor
open circuit/not calibrated
- Temperature sensor damaged
- Temperature sensor short circuit
- Temperature sensor missing
E2
Humidity sensor open
circuit/not calibrated
- Condensation on humidity sensor
- Humidity sensor damaged
- Humidity sensor short circuit
- Humidity sensor missing
E5
Capasitance out of range - Condensation on humidity sensor
- Humidity sensor damaged
E6
Program checksum error Internal error
E7 Parameter
storage
checksum error
Internal error
E9 Sensor
coefficient
checksum error
Internal error
E10 Current
settings
checksum error
Internal error
E11 Factory
settings
checksum error
Internal error
E12 Temperature
sensor
values out of range
- Temperature sensor damaged
- Temperature sensor missing
E14
MeasTask not calculating
(ATEST)
This error may occurr when the
analog output is in test mode. See
page Test Analog Output on page
42.
E16
Default settings not found Internal error
E17
RH reading too high
Internal error
E18
RH reading too low
Internal error
E20
Log corrupted
Internal error