beautypg.com

List of serial commands – Vaisala DMT132 User Manual

Page 38

background image

USER'S GUIDE____________________________________________________________________

36 __________________________________________________________________ M211289EN-A

List of Serial Commands

All commands can be issued either in uppercase or lowercase. In the
command examples, the keyboard input by the user is in bold type.

The notation refers to pressing the carriage return (Enter) key on
your computer keyboard. Enter a to clear the command buffer
before starting to enter commands.

Table 6

DMT132 Serial Commands

Command Description
? Show

device

information

AERR

Set analog output device malfunction error mode

ALARM

Set analog output limit alarm modes

ASEL

Configure analog output quantities and scaling

ATEST

Test analog outputs

CALCS

Show available quantities

CRH

Adjust capacitance measurement after sensor change

CT

Calibrate temperature measurement

ERRS

Display the currently active errors

FRESTORE

Restore factory settings

HELP

Display the command list

L

View user adjustment parameters

LI

Set user adjustment parameters

PASS

Access advanced service commands. Requires
passcode. Use when instructed by Vaisala personnel.

PRES

Set value for pressure compensation

RHCAL

Calibrate humidity measurement

RESET

Reset the transmitter

SEND

Output the reading once

SYSTEM Show

transmitter software information

VERS

Show transmitter firmware version