Serial line output commands, Output a reading once, Troubleshooting and maintenance commands – Vaisala DMT132 User Manual

Page 43: Display currently active errors

background image

Chapter 4 ________________________________________________________________ Operation

VAISALA _______________________________________________________________________ 41

NOTE

Conversions from mmHg and inHg are defined at 0 °C and from
mmH

2

O and inH

2

O at 4 °C.

NOTE

You can also set the pressure compensation value when you connect the
DMT132 to the MI70 indicator. When using the MI70, the pressure unit
is bar

a

.

Serial Line Output Commands

Output a Reading Once

Use the SEND command to output a single measurement message. The
output message is sent for the quantity that is currently selected using the
ASEL

command.

SEND

Example:

>send
Tdf= -0.80'C
>

Troubleshooting and Maintenance Commands

Display Currently Active Errors

Use the ERRS command to display the currently active transmitter error
codes. Possible errors and their causes are listed in Table 8 on page 65.

ERRS

Example (no errors active):

>errs
No errors
>

Example (Error E2 active):

>errs
Error: E2 Humidity sensor open circuit/not calibrated
>