Serial line output commands, Output a reading once, Troubleshooting and maintenance commands – Vaisala DMT132 User Manual
Page 43: Display currently active errors

Chapter 4 ________________________________________________________________ Operation
VAISALA _______________________________________________________________________ 41
NOTE
Conversions from mmHg and inHg are defined at 0 °C and from
mmH
2
O and inH
2
O at 4 °C.
NOTE
You can also set the pressure compensation value when you connect the
DMT132 to the MI70 indicator. When using the MI70, the pressure unit
is bar
a
.
Serial Line Output Commands
Output a Reading Once
Use the SEND command to output a single measurement message. The
output message is sent for the quantity that is currently selected using the
ASEL
command.
SEND
Example:
>send
Tdf= -0.80'C
>
Troubleshooting and Maintenance Commands
Display Currently Active Errors
Use the ERRS command to display the currently active transmitter error
codes. Possible errors and their causes are listed in Table 8 on page 65.
ERRS
Example (no errors active):
>errs
No errors
>
Example (Error E2 active):
>errs
Error: E2 Humidity sensor open circuit/not calibrated
>