beautypg.com

Technical data, Specifications, Performance – Vaisala DMT132 User Manual

Page 69: Measured variables, Dmt132, Chapter 7

background image

Chapter 7 ____________________________________________________________ Technical Data

VAISALA _______________________________________________________________________ 67

CHAPTER 7

TECHNICAL DATA

This chapter provides the technical data of the DMT132.

Specifications

Performance

Measured Variables

Dewpoint temperature
Measurement range

-20 ... +50 °C (-4 ... +122 °F) T

d

Accuracy at +20 °C (+68 °F)

±1 °C for -3 ... 20 °C (+26.6 ... +68 °F) T

d

±2 °C for -15 ... -3 °C (+5 ... +26.6 °F) T

d

Response time at 20 °C (+68 °F) gas temperature and 1 bar pressure

-20 -> +50 °C (-4 -> +122 °F) T

d

17 s (63 %)

40 s (90 %)

+50 -> -20 °C (+122 -> -4 °F)T

d

33 s (63 %)

85 s (90 %)

D

e

w

po

int

t

e

m

p

er

at

u

re

(

°C)

0

20

20

0

10

30

-10

-20

-30

-10

10

30

40

50

60

Temperature of measured gas (°C)

50

40

60

Accuracy

±1 °C

Accuracy

±2 °C

1101-058

Figure 28

Accuracy Over Temperature Range

When dewpoint is below 0 °C, the transmitter outputs frostpoint for T

d

.