USER'S GUIDE____________________________________________________________________

58 __________________________________________________________________ M211289EN-A

Temperature Calibration and Adjustment
(CT)

The procedure below describes a temperature (T) measurement
calibration and adjustment using serial line commands. You can use the
procedure to do a one-point calibration (offset) or a two-point calibration
(offset and gain). For two-point calibration, the second reference point
must be at least 20 ºC warmer than the first.

1.

Disconnect the cable from connector I of the DMT132 (if
connected).

2.

Connect the USB cable to connector II on the DMT132, and to a
USB port on your PC.

3.

Start a terminal application and set the service port connection
settings (19200 8 N 1). Remember to check which COM port the
USB cable is using.

4.

Verify that the connection works by giving the ? command:

>?


If the transmitter does not respond with device information:

- Check your serial line settings and cable connections

- Reconnect the USB cable, restart the terminal application, and

try again.

5.

When your serial connection is working, use the L command to see
the current user adjustment parameters. Write down the T offset
and gain values so you can restore them manually using the LI
command, if necessary.

>l

6. Use

the

ERRS

command to see that no errors are active:

>errs

7.

Insert the transmitter in the temperature reference.

8. Give

the

CT

command. The output from the command shows the

currently measured T value, and prompts you to enter the value of
the first reference (Ref1). This is the reference on the colder end,
which produces the offset correction for the measurement.

>ct
T 21.0947 Ref1 ?