beautypg.com

Vaisala HMT310 User Manual

Page 74

background image

USER'S GUIDE____________________________________________________________________

74 __________________________________________________________________ M210619EN-D

Table 12

Temperature Specifications (and Operating Pressure
Ranges)

Property

Description / Value

HMT311

-40 ... +60 °C (-40 ... +140 °F)

HMT313 80 °C

-40 ... +80 °C (-40 ... +176 °F)

HMT313 120 °C

-40 ... +120 °C (-40 ... +248 °F)

HMT314

-70 ... +180 °C (-94 ... +356 °F),
0 ... 10 MPa (0 ... 100 bar)

HMT315 (vapor tight)

-70 ... +180 °C (-94 ... +356 °F)

HMT317 (vapor tight)

-70 ... +180 °C (-94 ... +356 °F)

HMT318

-70 ... +180 °C (-94 ... +356 °F),
0 ... 4 MPa (0 ... 40 bar)

Accuracy at +20 °C (+68 °F)

±0.2 °C

Accuracy over temperature range

See Figure 18 below.

Typical temperature dependence of
electronics

±0.05°C/°C (±0.005°F/°F)

Temperature sensor

Pt 100 RTD 1/3 Class B IEC 751

0507-021

Figure 18

Accuracy over Temperature Range