Chapter 8 ____________________________________________________________ Technical Data

VAISALA _______________________________________________________________________ 73

CHAPTER 8

TECHNICAL DATA

This chapter provides the technical data of the product.

Specifications

Table 11

Relative Humidity Specifications

Property

Description / Value

Measuring range

0 ... 100 %RH

Accuracy (including nonlinearity,
hysteresis and repeatability)
with
Vaisala HUMICAP

®

180R

Vaisala HUMICAP

®

180RC


at +15 ... 25 °C

at -20 ... +40 °C
at -40 ... +180 °C




for typical applications
for applications with chemical purge
and/or warmed probe
± 1 % RH (0 ... 90 %RH)
± 1.7 % RH (90 ... 100 %RH)
± (1.0 + 0.008 × reading) %RH
± (1.5 + 0.015 × reading) %RH

Factory calibration uncertainty
(+20 °C)

±0.6 % RH (0 ... 40 %RH)
±1.0 % RH (40 ... 97 %RH)
(Defined as ± 2 standard deviation
limits. Small variations possible, see
also calibration certificate.)

Response time (90 %) for
HUMICAP

®

180R and

HUMICAP

®

180RC at 20 °C in

0.1 m/s air flow

17 s with grid filter
50 s with grid + steel netting filter
60 s with sintered filter