Ctext set calibration information text, Cdate set calibration date – Vaisala HMT310 User Manual
Page 68

USER'S GUIDE____________________________________________________________________
68 __________________________________________________________________ M210619EN-D
FCRH Relative Humidity Calibration after
Sensor Change
Syntax: FCRH
The transmitter asks and measures relative humidity and calculates the
calibration coefficients. This two-point adjustment needs to be performed
after a sensor change. Follow the more detailed calibration instructions
on page 69, but instead of the CRH command, use the FRCH command.
Example:
>frch
RH :
1.82
1.
ref
?
0
Press any key when ready
RH
:
74.22
2. ref
? 75
OK
>
The OK indicates that the calibration has succeeded.
CTEXT Set Calibration Information Text
Syntax: CTEXT
1.
Remove the transmitter unit from the mounting plate (see
Mounting the Transmitter/Removing the Transmitter Unit on page
21) and press the adjustment button once (see Figure 17 on page
70).
2.
Issue command LI and enter value 0 for an offset value and value 1
for a gain value.
3.
Issue command CTEXT and enter calibration information text.
Example:
>ctext
Cal. info : Vaisala/HEL ? HMK15
>
CDATE Set Calibration Date
Syntax: CDATE yyyy mm dd
1.
Remove the transmitter unit from the mounting plate (see
Mounting the Transmitter/Removing the Transmitter Unit on page
21) and press the adjustment button once (see Figure 17 on page
70).