USER'S GUIDE____________________________________________________________________

68 __________________________________________________________________ M210619EN-D

FCRH Relative Humidity Calibration after
Sensor Change

Syntax: FCRH

The transmitter asks and measures relative humidity and calculates the
calibration coefficients. This two-point adjustment needs to be performed
after a sensor change. Follow the more detailed calibration instructions
on page 69, but instead of the CRH command, use the FRCH command.

Example:

>frch
RH :

1.82

1.

ref

?

0

Press any key when ready
RH

:

74.22

2. ref

? 75

OK
>

The OK indicates that the calibration has succeeded.

CTEXT Set Calibration Information Text

Syntax: CTEXT

1.

Remove the transmitter unit from the mounting plate (see
Mounting the Transmitter/Removing the Transmitter Unit on page
21) and press the adjustm
ent button once (see Figure 17 on page
70).

2.

Issue command LI and enter value 0 for an offset value and value 1
for a gain value.

3.

Issue command CTEXT and enter calibration information text.

Example:

>ctext
Cal. info : Vaisala/HEL ? HMK15
>

CDATE Set Calibration Date

Syntax: CDATE yyyy mm dd

1.

Remove the transmitter unit from the mounting plate (see
Mounting the Transmitter/Removing the Transmitter Unit on page
21) and press the adjustm
ent button once (see Figure 17 on page
70).