Itest test analog outputs, Aqtest test analog outputs for desired readings – Vaisala HMT310 User Manual
Page 53
Chapter 4 ________________________________________________________________ Operation
VAISALA _______________________________________________________________________ 53
ITEST Test Analog Outputs
Syntax: ITEST aa.aaa bb.bbb
The operation of the analog outputs are tested by forcing the outputs to
given values. The values in the analog outputs can then be measured with
a current/voltage meter.
Where
aa.aaa
=
Current value to be set for channel 1
bb.bbb
=
Current value to be set for channel 2
Example:
>itest 8
12
8.000mA
3F8 12.000mA 70O
>
The set values remain valid until you issue the ITEST command without
readings or reset the transmitter.
AQTEST Test Analog Outputs for
Desired Readings
Syntax: AQTEST x yyy.yyy
Use command AQTEST to test current values. Current output is forced to
correspond to the chosen values.
Where
x
=
Output quantity of analog channel (use
abbreviations, see FORM Serial Output Format on
page 43)
yyy.yyy =
Value
Example:
>aqtest td 30
CH1 Td : 30.0000 'C 12.00000mA
CH2 x : 2.5304 g/kg 4.08097mA
>
The set values remain valid until you give the command AQTEST
without a value or reset the transmitter.