Hmt315 for high temperatures, Hmt317 for demanding processes – Vaisala HMT310 User Manual

Page 28

background image

USER'S GUIDE____________________________________________________________________

28 __________________________________________________________________ M210619EN-D

NOTE

When HMT314 is installed in a process with pressure differing from
normal atmospheric pressure, please enter the pressure value of the
process (in hPa or mbar) into the transmitter memory via the serial line
(see PRES Set Ambient Pressure for Calculations on page 48).

HMT315 for High Temperatures

HMT315 is installed similarly than the HMT313 probe but without the
supporting bar. Refer to Appendix A on page 81 for more information on
the duct installation kit for HMT315.

To avoid incorrect humidity readings, the temperature differences
between inside and outside of the duct must not be remarkable.

HMT317 for Demanding Processes

The HMT317 is for environment where relative humidity is very high,
near saturation. The warmed probe prevents the saturation of the sensor.

See Appendix A on page 81 for a presentation of the following probe
installation kits for HMT317 with installation examples:

- Duct mounting kit

- Cable

gland

- Pressure tight Swagelok connector