USER'S GUIDE____________________________________________________________________

64 __________________________________________________________________ M210619EN-D

1.

Remove the filter from the probe.

2.

Remove the damaged sensor and insert a new one. Handle the new
sensor by the plastic socket.

CAUTION

Do not touch the sensor plate.

3.

After sensor change the humidity calibration must be made
according to the instructions, see FCRH Relative Humidity
Calibration after Sensor Change on page 68.

4.

Install a new filter on the probe. When using the stainless steel
filter, take care to tighten the filter properly (recommended force is
130 Ncm).

New sensors can be ordered from Vaisala, see Parts List for
Consumables below.

Parts List for Consumables

Table 10

Available Spare Parts

Spare Part

Order Code

PPS Plastic Grid with Stainless Steel Netting DRW010281SP
PPS Plastic Grid Filter

DRW010276SP

Sintered Filter AISI 316L

HM47280SP

Stainless Steel Filter

HM47453SP

Stainless Steel Filter with Membrane

214848SP

Vaisala HUMICAP

®

180R (general purpose)

HUMICAP180R

PT100 sensor

10429SP

Technical Support

For technical questions, contact the Vaisala technical support:

E-mail

[email protected]

Fax

+358 9 8949 2790