1.
Remove the filter from the probe.
2.
Remove the damaged sensor and insert a new one. Handle the new
sensor by the plastic socket.
CAUTION
Do not touch the sensor plate.
3.
After sensor change the humidity calibration must be made
according to the instructions, see FCRH Relative Humidity
Calibration after Sensor Change on page 68.
4.
Install a new filter on the probe. When using the stainless steel
filter, take care to tighten the filter properly (recommended force is
130 Ncm).
New sensors can be ordered from Vaisala, see Parts List for
Consumables below.
Parts List for Consumables
Table 10
Available Spare Parts
Spare Part
Order Code
PPS Plastic Grid with Stainless Steel Netting DRW010281SP
PPS Plastic Grid Filter
DRW010276SP
Sintered Filter AISI 316L
HM47280SP
Stainless Steel Filter
HM47453SP
Stainless Steel Filter with Membrane
214848SP
Vaisala HUMICAP
®
180R (general purpose)
HUMICAP180R
PT100 sensor
10429SP