USER'S GUIDE____________________________________________________________________

46 __________________________________________________________________ M210619EN-D

Table 7

Output Quantities and their Metric and Nonmetric
Units

Quantity

Metric Unit

Nonmetric Unit

RH Relative Humidity

%RH

%RH

T Temperature

C

F

TDF Dewpoint/frostpoint

C

F

TD Dewpoint

C

F

A Absolute humidity

g/m

3

gr/ft

3

X Mixing ratio

g/kg

gr/lb

TW Wet bulb temperature

C

F

PPM Humid air volume/dry air volume

Ppm

ppm

PW Water vapor pressure

hPa

lb/in

2

PWS Saturation water vapor pressure

hPa

lb/in

2

H Enthalpy

kJ/kg

Btu/lb

Example:

>unit m
Output units : metric
>unit n
Output units : non metric
>

FST Output State of Chemical Purge or
Sensor Heating (with SEND and R
Commands)

Syntax: FST x

Where

x = ON/OFF (default = OFF)

Example:

>fst on
Form. status : ON
>send

1

N 0 RH= 40.1 %RH T= 24.0 'C Td= 9.7 'C Tdf= 9.7 'C a=

8.7 g/m3 x= 7.5
g/kg Tw= 15.6 'C ppm= 11980 pw= 12.00 hPa pws= 29.91 hPa
h= 43.2 kJ/kg

>purge
Purge started, press any key to abort.

>send

1

S 134 RH= 40.2 %RH T= 24.1 'C Td= 9.8 'C Tdf= 9.8 'C a=

8.8 g/m3 x= 7.5