Chapter 4 ________________________________________________________________ Operation

VAISALA _______________________________________________________________________ 49

Example:

If gauge pressure is 1.4 bar, set pressure value 2.4 bar (= 2400 hPa)

>pres 2400
Pressure :

2400

>

Table 8

Pressure Conversion Chart

TO

FROM

hPa
mbar

mmHg
Torr

inHg atm

bar

psi

hPa

mbar

1 1.333224

33.86388

1013.25

1000

68.94757

mmHg
Torr

0.7500617 1

25.40000 760

750.0617

51.71493

inHg

0.02952999 0.03937008 1

29.921 29.52999 2.036021

atm

0.00098692 0.00131597 0.033422

1

0.98692 0.068046

bar

0.001 0.001333224 0.03386388

1.01325

1 0.06894757

psi

0.01450377 0.01933678 0.4911541 14.6962 14.50377 1

Example:

29.9213 inHg = 29.9213 x 33.86388 = 1013.25 hPa/mbar

NOTE

Conversions from mmHg and inHg are defined at 0 °C and for mmH2O
and inH2O at 4 °C.

NOTE

Pressure compensation is intended to be used in normal air only. When