Mixing ratio, Absolute humidity, Enthalpy – Vaisala HMT130 User Manual
Page 89

Appendix A _______________________________________________________ Calculation Formulas
VAISALA ________________________________________________________________________ 87
Mixing Ratio
The mixing ratio (mass of water vapour/mass of dry gas) is calculated
using:
w
w
p-P
P
B
x
×
=
[
]
kg
g /
(3)
where
B = 621.9907 g/kg
The value of B depends on the gas. 621.9907 g/kg is valid for air.
Absolute Humidity
Absoute humidity is defined as the mass of water vapour in a certain
volume. If ideal gas behaviour is assumed the absolute humidity can be
calculated using:
(
)
15
.
273
+
⋅
=
T
P
C
a
w
[
]
3
/ m
g
(4)
where
C = 216.679 gK/J
Enthalpy
Entalpy can be calculated from mixing ratio using:
x
x
T
h
⋅
+
⋅
+
⋅
=
5
.
2
)
00189
.
0
01
.
1
(
[
]
kg
kJ /
(5)
See also other documents in the category Vaisala Humidifiers:
- Calibration of Digital Transmitters with HMI41 (36 pages)
- Calibration of Series HMDW2030 and HMP130 Transmitter with HMI41 (14 pages)
- Calibration of Series HMDW6070 and HMP140 Transmitter with HMI41 (30 pages)
- HM34 (30 pages)
- HM40 (47 pages)
- HM44 (52 pages)
- HM70 (83 pages)
- HMD40 (1 page)
- HMD60 (4 pages)
- HMD70 (18 pages)
- HMDW110 (62 pages)
- HMDW80 (51 pages)
- HMI41 (74 pages)
- HMP41 (72 pages)
- HMK15 (39 pages)
- HMM100 (71 pages)
- HMM105 (23 pages)
- HMM211 (42 pages)
- HMM212 (36 pages)
- HMM213 (52 pages)
- HMP140 (28 pages)
- HMP155 (84 pages)
- HMP228 (115 pages)
- HMP230 (163 pages)
- HMP240 (130 pages)
- HMP260 (118 pages)
- HMP60 (71 pages)
- HMT100 (52 pages)
- HMT120 (87 pages)
- HMT140 (76 pages)
- HMT310 (88 pages)
- HMT310 (105 pages)
- HMT330 (209 pages)
- HMT360 (97 pages)
- HMT360 (63 pages)
- HMT360N (110 pages)
- HMW40 (1 page)
- HMW90 (110 pages)
- SHM40 (68 pages)
- RDP100 (14 pages)