Chapter 4 ________________________________________________________ Serial Line Operation

VAISALA ________________________________________________________________________ 31

Device Information and Status

Show Device Information

The ? command outputs a listing of device information.

?

Example:

>?

Device Name : HMT130

SW Name : HMT120/130

SW model : HMT130

SW version : 0.9.3.389

Serial number : "F2220101"

Address : 0

Unit : METRIC

Ch1 Status : ON

Ch2 Status : ON

Probe name : "HMP110"

Probe SW : "1.01.1"

Probe SW date : ""

Probe SN : "F0740011"

>

Set/Show Available Quantities

The CALCS command shows the abbreviations of the quantities

measured by HMT130. The command can also be used to select the

desired quantities for measurement. Only after the quantities have been

set with the CALCS command, can these quantities be selected for

different outputs (see commands DSEL, ASEL, and FORM). The

quantities in DSEL, ASEL, RSEL, and FORM must match with

CALCS.

CALCS [q1 q2]

Where
q1, q2 = Any two of the allowed quantity parameters, i.e. RH, T,

TD, TDF, A, X, H, TW, PWS, PW

Examples:

>calcs

RH Td

>

calcs td t

>