beautypg.com

Set/show analog output error levels, Set/show analog output quantity and scaling – Vaisala HMT130 User Manual

Page 44

background image

User's Guide _______________________________________________________________________

42 ___________________________________________________________________ M211280EN-B

Set/Show Analog Output Error Levels

Use the AERR command to show or set the analog output error levels.

AERR [level1 level2]

where
level1 level2 = Error levels for analog output channel

Examples:

>aerr

Ch1 Error level : 0.0

Ch2 Error level : 0.0

>

>aerr 10.8 10.8

Ch1 Error level : 10.8

Ch2 Error level : 10.8

>

Set/Show Analog Output Quantity and

Scaling

After you have defined the quantities with the CALCS command, use the

ASEL command to show or set the quantity output by the analog

channels, and the scaling that is used. If you do not enter scaling limits,

default values are used. Valid quantities are: RH, T, TD, TDF, A, X, H,

TW, PWS, PW and NONE. NONE parameter disables analog output and

output is forced to error level.

ASEL [q1 q2] [lo1 hi1 lo2 hi2]

where
q1 q2

= Analog output channel quantities

lo1 lo2 = Low limits of scaling

hi1 hi2 = High limits of scaling
Syntax for single-parameter transmitters:

ASEL [q] [lo hi]

Where
q

= Analog output channel quantity

lo

= Low limit of scaling

hi

= High limit of scaling