Test analog outputs, Configuring relay output, Set/show relay quantity – Vaisala HMT130 User Manual

Page 46

background image

User's Guide _______________________________________________________________________

44 ___________________________________________________________________ M211280EN-B

Test Analog Outputs

Use the ATEST command to test the analog outputs. The ATEST

command will force the output to the given value, which can then be

measured with a calibrated multimeter.

ATEST command affects only the analog outputs and hence it does not

disable measurement functions. ATEST without parameters will exit the

test mode and let the measurement continue. It will also indicate the the

current analog output levels in normal measurement state.

ATEST [val1 val2]

where
val1 val2

= Analog channel output value (V)

Examples:

>atest

CH1: 2.360000

CH2: 3.765010

>

>atest 5.0 5.0

CH1: 5.000000

CH2: 5.000000

>

Configuring Relay Output

Set/Show Relay Quantity

After you have defined the quantities with the CALCS command, use the

RSEL command to show or set the relay quantity. Valid quantity

parameters are RH, T, TD, TDF, A, X, H, TW, PWS and PW.

RSEL [q]

where
q

= Quantity to define relay state.

Example:

>RSEL T

Relay quantity : T