beautypg.com

Temperature field check and adjustment by using a, Calibrated reference probe – Vaisala HMT130 User Manual

Page 68

background image

User's Guide _______________________________________________________________________

66 ___________________________________________________________________ M211280EN-B

Temperature Field Check and Adjustment by Using a

Calibrated Reference Probe

Follow steps 1 to 7 on page 62 and continue as follows:

8.

Check that the probes are located in equal conditions and wait

until the readings are stabilized (can take 30 minutes or more). If

you are close to the probes, do not breathe in their direction.

9.

Press

ADJUST

to continue adjusting.

10. Press

To same as T

II/I

, press

SELECT

. (MI70 always recognizes the

port to which the HMP70 series probe is connected).

11. Confirm by pressing

YES

.

NOTE

If the temperature difference between the reference probe and HMT130

is too large, the adjustment cannot be done (HM70 will notify you of

this). The available reserve for T adjustment of HMT130 depends on

the initial temperature calibration of the unit.

12. Adjustment is done. Press

BACK

and

EXIT

to return to the basic

display.

13. Switch off the MI70 and detach the calibration cable from MI70

and from the transmitter.