10. Deflate the bulb by attaching an appropriately sized luer lock sy-

ringe to the valve, engage the valve and slowly depress the bulb

allowing the syringe to fill with air. Alternatively, remove plunger

from the syringe, attach syringe and allow air to slowly release

while gently depressing the bulb.

Note: Do not draw negative pressure in the syringe, as this will create

a vacuum on the site.

11. Prior to discharge of the patient, remove SAFEGUARD and apply

sterile dressing per hospital protocol.

POST-HEMOSTASIS TECHNIQUE

1. When hemostasis at the access site has

been achieved, apply the SAFEGUARD

device with the access site visible under

the bulb window of the SAFEGUARD

device. Consider the point of maximum

pulse, anatomy, angle of puncture

and direction of flow to determine the

appropriate SAFEGUARD position and

verify.

Note: Before adhering SAFEGUARD to the

patient, be sure skin is clean and dry. Deter-

mine the appropriate angle for SAFEGUARD

placement to provide easy access to luer

inflate/deflate port.

2. Attach an appropriately sized standard

luer lock syringe to the valve of the

SAFEGUARD device.

Note: Syringe must be completely engaged

in the luer to inflate/deflate the bulb.

3. Inflate the bulb of the SAFEGUARD

device with air to the desired volume of

air (12 cm maximum of 7 mL’s) to apply

pressure on the arteriotomy site and re-

move the syringe. Check distal/proximal

pulses to assure flow is maintained.

Note: Maintain pressure on the plunger

while detaching syringe from the valve of

the SAFEGUARD device. Observe that the

desired volume is achieved and maintained.

4. Per hospital protocol, periodically check the site through the

bulb window to assure hemostasis is maintained and the bulb

maintains pressure.

5. Deflate bulb every two hours and assess the site. Re-inflate the

bulb if necessary.

6. Deflate the bulb by attaching an appropriately sized luer lock sy-

ringe to the valve, engage the valve and slowly depress the bulb

allowing the syringe to fill with air. Alternatively, remove plunger

from the syringe, attach syringe and allow air to slowly release

while gently depressing the bulb.

Note: Do not draw negative pressure in the syringe, as this will create

a vacuum on the site.

7. Prior to discharge of the patient, remove the SAFEGUARD device

and apply sterile dressing per hospital protocol.

Maintain sterile field during application.

Keep away from sunlight

Do not Resterilize

Do not Reuse

Do not use if package is damaged

Catalog Number

Lot Number

Use By

Sterilized Using Gamma

Caution: Consult accompanying document

Manufacturer

Authorized Representative

DISPOSITIF DE 12 cm ASSISTÉ PAR PRESSION

MODE D’EMPLOI

Lisez les instructions avant utilisation.

Produit dépourvu de caoutchouc de latex naturel.

DESCRIPTION DU DISPOSITIF

Le SAFEGUARD de12 cm est un dispositif jetable à usage unique.

Le SAFEGUARD possède une fenêtre et une vessie transparentes en

polyuréthane de qualité médicale, un tube de remplissage en PVC

transparent et souple de qualité médicale, et un support auto-adhésif

sensible à la pression. Une valve Luer à l’extrémité du tube de

remplissage permet de connecter une seringue pour gonfler d’air la

vessie centrale afin de fournir une pression au site de la ponction. Le

dispositif SAFEGUARD assisté par pression contient un pansement

stérile avec une fenêtre transparente qui facilite la visibilité du site

d’accès sans devoir retirer ou manipuler l’appareil.

INDICATIONS

L’utilisation de l’appareil SAFEGUARD assisté par pression de 12 cm

est indiquée pour obtenir et maintenir l’hémostase.

CONTRE-INDICATIONS

La partie adhésive du dispositif SAFEGUARD ne doit pas être utilisée

sur une peau présentant des excoriations.

MISES EN GARDE

Px Only. : Mise en garde : En vertu de la loi fédérale (États-Unis), ce

dispositif est strictement réservé à la vente par un médecin autorisé,

ou sur son ordre.

MISES EN GARDE : Lors de sur-inflation, c’est-à-dire au-delà de 40

ml d’air, le ballonnet peut commencer à se dilater radialement et

pourrait compromettre les propriétés adhésives de l’appareil.

MISES EN GARDE :Une sous-inflation du dispositif pourrait compro-

mettre la capacité de l’appareil à obtenir et à maintenir l’hémostase.