Merit Medical ONE Snare Endovascular Snare System User Manual

Page 3

background image

DESCRIPTION: English

The ONE Snare™ endovascular snare system consists of the snare, snare catheter, insertion tool and

torque. The snare is constructed of nitinol cable and gold plated tungsten loop. The pre-formed

snare loop can be introduced through catheters without risk of snare deformation because of the

snare’s super-elastic construction. The snare catheter is constructed of polyether block amide and

contains a platinum/iridium radiopaque marker band.

INDICATION FOR USE:

The ONE Snare™ endovascular snare system is intended for use in the coronary and peripheral vascular

system or hollow viscous to retrieve and manipulate foreign objects. Retrieval and manipulation

procedures include indwelling venous catheter repositioning, indwelling venous catheter fibrin sheath

stripping, and central venous access venipuncture procedure assistance.

CONTRAINDICATIONS:

1. This device is not intended for the removal of foreign objects entrapped by tissue growth.

2. This device should not be used for fibrin sheath stripping in the presence of septal defects

of Persistent Foramen Ovale.

3. This device is not intended for removal of implanted pacing leads.

WARNINGS:

1. Pull forces applied to catheters during fibrin sheath stripping may damage, stretch, or break

indwelling catheters 6 French or smaller in diameter. Do not use excessive pull force when

attempting fibrin sheath stripping of catheters 6 French or smaller in diameter.

2. Do not use excessive force when manipulating the catheter through an introducer.

Excessive force may damage the snare catheter.

3. This device has been sterilized utilizing ethylene oxide and is considered sterile if the

package is not opened or damaged. Do not use a device that has been damaged or if the

package is open or damaged.

4. For single patient use only. Do not reuse, reprocess or sterilize. Reuse, reprocessing or

resterilization may compromise the structural integrity of the device and/or lead to device

failure which, in turn, may result in patient injury, illness or death. Reuse, reprocessing,

or resterilization may also create a risk of contamination of the device and/or cause

patient infection or cross-infection including, but not limited to, the transmission of

infectious disease(s) from one patient to another. Contamination of the device may lead

to injury, illness or death of the patient.

5. After use this device may be a potential biohazard. Handle in a manner that will

prevent accidental contamination.

6. Nitinol is a nickel titanium alloy. Possible reaction may occur for those patients who

exhibit sensitivity to nickel.

7. Care should be observed when using this device for removal of a large fibrin sheath in order to

minimize risk of pulmonary embolism.

PX only: CAUTION - Federal (U.S.A.) law restricts this device to sale by or on the order of a physician.
POTENTIAL COMPLICATIONS:

1. Potential complications associated with foreign body retrieval devices in arterial

vasculature include, but are not limited to:

• Embolization

• Stroke

• Myocardial infarction (depending upon placement)

2. Potential complications associated with snare retrieval devices in venous vasculature

include, but are not limited to:

• Pulmonary embolism

3. Other potential complications associated with foreign body retrieval devices include,

but are not limited to:

• Vessel perforation

• Device entrapment

Catheter damage can occur when attempting fibrin sheath stripping on small French size diam-

eter catheters. (See WARNINGS)

Incidence of pulmonary embolism after fibrin sheath stripping may occur.

(See WARNINGS)

DIRECTIONS FOR USE:

Prepare the ONE Snare™ System:

Select the appropriate snare diameter range for the site in which the foreign body is located. The snare

diameter range should approximate the size of the vessel in which it will be used.

1. Remove the snare and snare catheter from their hoop holders and inspect for any damage.

2. Remove the insertion tool and torque device from the proximal end of the snare shaft.

3. Load the snare into the snare catheter by inserting the proximal end of the snare into the distal

(non-hubbed) end of the snare catheter, until the proximal end of the snare shaft exits the hub and

the loop can be retracted into the distal end of the snare catheter.

4. Inspect the device by extending and retracting the snare loop through the distal end of the snare

catheter 2-3 times, while carefully examining the snare catheter, radiopaque band and the device

for any damage or defects.

5. When appropriate, the system (snare and snare catheter) can be advanced to the desired site as a

single unit assembled as described above.

Alternative Preparation of the ONE Snare™ System:

If the snare catheter is already positioned within the vasculature, the provided insertion tool (located

on the proximal end of the snare and just distal to the torque device) may be used to position the snare

in the indwelling snare catheter.

1. Remove the snare from the protective holder and inspect for any damage.

2. Move the provided insertion tool (located on the proximal end of the snare, just distal to the torque

device) distally until the loop of the snare is enclosed within the tubing portion of the insertion tool.

3. Insert the distal end of the insertion tool into the hub of the indwelling snare catheter until

resistance is felt. This will indicate the tip of the insertion tool is properly aligned with the inner

lumen.

4. Hold the insertion tool as straight as possible, grasp the shaft of the snare just proximal to the hub

of the insertion tool and advance the snare until it is secure within the lumen of the snare catheter.

The insertion tool can be removed by grasping the blue tab and firmly peeling it away from the

snare shaft.

Snare Assisted Retrieval and Manipulation Suggestions:

1. If present, remove the indwelling delivery catheter.

2. If a guidewire is in a patient at the location of a foreign body, advance a snare catheter over the

guidewire to the desired location. Then remove the guidewire and advance the snare through the

snare catheter. Alternatively, cinch one loop of the snare over the proximal end of the guidewire and

advance the entire system (snare and snare catheter assembly) into a guide catheter or introducer

sheath until the distal end of the snare catheter is positioned proximal to the foreign body.

3. If a guidewire is not present, pull the snare into the distal end of the snare catheter and advance

through a guide catheter or introducer sheath until it is positioned proximal to the foreign body.

Alternatively, collapse the snare loop by pulling the device into the distal end of the snare insertion

tool. Place the tapered end of the snare insertion tool into the proximal (hub) end of the snare

catheter, guide catheter or sheath and advance the snare forward maintaining constant contact

between the insertion tool and snare catheter hub. NOTE: When attempting to utilize guide

catheters or sheaths not specifically manufactured for use with the ONE Snare™ system, it is

important to test product compatibility prior to use.

4. Gently push the snare shaft forward to completely open the loop. The loop is then slowly advanced

forward, and around the proximal end of the foreign body. Alternatively, the snare may be advanced

beyond the target location and the loops brought back around the distal end of the foreign body.

5. By advancing the snare catheter, the loop of the snare is closed to capture the foreign body. Note

that attempting to close the loop by pulling the snare into the snare catheter will move the loop

from its position around the foreign body.

6. To manipulate a foreign body, maintain tension on the snare catheter to retain the hold on the

foreign body, and move the snare and snare catheter together to manipulate a foreign body to the

desired position.

7. To retrieve a foreign body, maintain tension on the snare catheter and move the snare and snare

catheter assembly together proximally to or into a guide catheter or sheath. The foreign body is

then withdrawn through or together with the guiding catheter or introducer sheath. Withdrawal of

large foreign bodies may require the insertion of larger sheaths, guiding catheters, or a cut-down at

the peripheral site.

Snare Assisted Removal of Fibrin Sheaths from Indwelling Catheters:

1. Using standard technique, prepare a femoral vein approach, advance the selected snare to the

inferior vena cava or right atrium.

2. Advance a .035” (0.89 mm) guidewire through the end port (distal or venous port if more than one

lumen) of the indwelling catheter and into the inferior vena cava or right atrium.

3. Position the snare loop around the guidewire.

4. Advance the snare over the distal end of the catheter to a position proximal to the fibrin sheath.

5. Close the snare around the catheter and continue applying light traction while gently pulling the

snare down toward the distal end of the catheter over the end ports.

6. Repeat steps 4 & 5 until the catheter is free of fibrin sheath.

Snare Assisted Venous Canalization:

1. Introduce the snare at a patent venous access site and position in the vasculature at the desired site.

2. Open the snare loop to provide a target to guide an entry needle into the desired venous access site.

3. Introduce a guidewire through the needle and through the snare loop. Remove the needle.

4. Close the snare over the guidewire by advancing the snare catheter.

5. Pull the guidewire into the inferior vena cava.

ONE Snare™ System contains:

(1) Snare, (1) Snare Catheter, (1) Insertion Tool and (1) Torque Device.

Merit Medical Systems, Inc.

1600 West Merit Parkway, South Jordan, Utah 84095 U.S.A. 1-801-253-1600

U.S.A. Customer Service 1-800-356-3748

Merit Medical Ireland Ltd, Parkmore Business Park West, Galway, Ireland

www.merit.com

I N S T R U C T I O N S F O R U S E :