Overview
Perform the following sequence in order when setting up
the table:
- Uncrating
-
Leveling the Base
-
Installing Grab Bars
Uncrating
NOTE: Inspect all boxes and contents for damage. Re-
port any damage to the carrier immediately.
CAUTION
To avoid damaging the table’s upholstery or paint-
ed surfaces, DO NOT use a knife or other sharp
object to open the packaging. Also, to avoid dam-
aging the table lift only at points indicated in Figure
3. Do not lift at other points as indicated by the
note in Figure 4.
CAUTION
Do not plug power into table prior to removal of all
packaging materials.
1. Remove stretch wrap surrounding table.
2. Cut and remove banding and protective foam tube
attaching table to pallet.
3. Remove the 6 steel shipping brackets securing the table
to the pallet, one at each of the four corners and one
at the middle on each side of the table (Figure 3). The
brackets engage with the table levelers to hold the table
securely to the pallet. Remove the 2 bolts from each
bracket using a 7/16” socket or wrench. Take care not
to scratch the unit when turning the wrench. Remove
brackets from levelers once bolts are removed.
4. Remove the 2 screws attaching the spacers on either
side of the wooden U-shaped brace located under the
front corners of the cabinet (Figure 4). Remove the
2 screws attaching the U-shaped brace to the pallet
(Figure 4).
5. Remove the foot control from the rear drawer. Insert the
foot control cord into the receptacle at the bottom of the
rear panel (
Figure 10)
. Secure with locking ring.
INSTALLATION
WARNING
The table weighs approximately 395 lbs. Use
enough personnel to remove the table from the
shipping pallet. Also, use proper lifting techniques
when lifting. Failure to do so could result in seri-
ous injury.
CAUTION
Do not use the stirrups to lift or move the table.
The Leg Support may be used to lift the table. Do
not pull on the Leg Support to slide the table.
See
SPECIFICATIONS
table for electrical rating of
this unit. The three-pronged grounding plug on the table
power cord must be plugged into a matching three-
pronged, grounded, non-isolated, correctly polarized 115
VAC receptacle.
CAUTION
Do not lift table by seat upholstery. Damage to
the pelvic tilt mechanism and/or upholstery may
occur.
Figure 3. Remove Shipping Brackets
Figure 4. Remove Spacers and U-Brace